TJ-II:Interferometry: Difference between revisions
Jump to navigation
Jump to search
mNo edit summary |
No edit summary |
||
Line 12: | Line 12: | ||
== The CO<sub>2</sub> interferometer == | == The CO<sub>2</sub> interferometer == | ||
The probing wavelength is about 10 μm. | The probing wavelength is about 10 μm. The system is located in [[TJ-II:Sectors|sector]] B1. |
Revision as of 10:43, 5 August 2009
TJ-II has two interferometry systems for the measurement of the line-integrated density.
The Microwave interferometer
The diagnostic is located in sector B8. The probing beam has an inclination of 18.7ΒΊ with respect to the vertical and a frequency of 140 GHz, corresponding to a wavelength of λ = 2.14 mm. The diagnostic provides a time trace of the-line integrated electron density (65536 points, sampling rate 100 kHz). The time trace is called 'Densidad2_' in the TJ-II database. The signal is less reliable for high densities (above 2-3 1019 m-3). This time trace is fundamental for machine operation.
The CO2 interferometer
The probing wavelength is about 10 μm. The system is located in sector B1.