TJ-II:Interferometry: Difference between revisions

From FusionWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 12: Line 12:
== The CO<sub>2</sub> interferometer ==
== The CO<sub>2</sub> interferometer ==


The probing wavelength is about 10 &mu;m.
The probing wavelength is 10.6 &mu;m.
<ref>[http://link.aip.org/link/?RSINAK/72/96/1 H. Lamela et al,  Rev. Sci. Instrum. '''72''', 96 (2001)]</ref>
<ref>[http://link.aip.org/link/?RSINAK/72/96/1 H. Lamela et al,  Rev. Sci. Instrum. '''72''', 96 (2001)]</ref>
The system is located in [[TJ-II:Sectors|sector]] B1.
The system is located in [[TJ-II:Sectors|sector]] B1.

Revision as of 17:57, 5 August 2009

TJ-II has two interferometry systems for the measurement of the line-integrated density.

The Microwave interferometer

The diagnostic is located in sector B8. The probing beam has an inclination of 18.7º with respect to the vertical and a frequency of 140 GHz, corresponding to a wavelength of λ = 2.14 mm. The diagnostic provides a time trace of the-line integrated electron density (65536 points, sampling rate 100 kHz). The time trace is called 'Densidad2_' in the TJ-II database. The signal is less reliable for high densities (above 2-3 1019 m-3). This time trace is fundamental for machine operation.

The CO2 interferometer

The probing wavelength is 10.6 μm. [1] The system is located in sector B1.

References